Source strength for high-dose-rate (HDR) and low-dose-rate
(LDR) brachytherapy sources should be measured by the clinical medical
physicist. In general, the manufacturer provides a calibration certificate
documenting the reference air-kerma rate (RAKR) or air-kerma strength (SK).
This certificate will be at a specific time point and for individual sources or
for the mean of a batch of sources. The physicist is responsibility for brachytherapy
dose calculations, which require input of source strengths, so physicist-measured
values are often used in the clinical process. Requirements for
physicist-measured values vary based on local regulatory standards, but these
measurements generally have lower uncertainties than manufacturer-reported
values. This is because the manufacturer transfers the calibration of a
traceably-calibrated instrument to the production instrument used for
calibrating customer sources while the chamber used by the medical physicist is
calibrated directly by the calibration laboratory.
The equipment needed to measure
brachytherapy source strength has become more uniform over time, and includes a
well-type air ionization chamber with an insert to reproducibly position the
source at the center of the collecting volume, a radiation source to
demonstrate system constancy, an electrometer and triaxial cabling, and a
barometer/thermometer/hygrometer (often all part of a single instrument). The
electrometer and triaxial cabling are typically shared with the dosimetry program
for external-beam radiotherapy calibrations. The radiation source to
demonstrate system constancy may be a long-lived radionuclide (e.g., 137Cs,
90Sr, or 241Am) having similar radiation quality and
source strength as the brachytherapy source to be measured. Other constancy sources
could include another brachytherapy (such as demonstrating system constancy for
a new HDR source by measuring an old HDR source preceding source exchange), or
even a linac with the well chamber positioned on the floor. The well chamber
and electrometer are sent for calibration every two years, or shown to have
constancy if intercompared with other instruments that have been calibrated
within two years. When mailing equipment for calibrations, the physicist should
measure before and afterwards that the instruments have not changed over the
course of shipment to demonstrate constancy of the calibration. All
calibrations should be directly traceable to a primary standards laboratory and
documented as such.
There are several guidance documents
and national laws on how to interpret physicist-measured values of
brachytherapy source strength in comparison to manufacturer-reported values.
For a single HDR source such as 192Ir or 60Co, the
measurement is compared and the physicist may use their value if the two agree
within 3%. For LDR seeds (125I, 103Pd, 131Cs),
often there is more than a single source and guidance documents generally
recommended that a subset be measured. When using stranded and sterile seeds to
implant, additional loose and non-sterile seeds (5% or 5 seeds, whichever is
fewer) from the same manufactured lot are ordered for the purpose of
demonstrating agreement within 3%. When comparison of the manufacturer and
physicist-measured values are not within the expected tolerance, the physicist
must investigate the source of discrepancy and alert the manufacturer (Butler et
al. MedPhys 2008). Sources with long half-lives (LDR 137Cs tubes
and HDR 90Sr/90Y ophthalmic applicators) are assayed at
least annually to show predictable decay.
The formalism for deriving
brachytherapy source strength is the product of the corrected reading,
multiplied by the well chamber calibration coefficient for the particular
source model, and decayed based on the source half-life to the reference day
and time. The reading is averaged over at least three separate measurements,
and corrections are made for ion recombination, atmospheric pressure and
temperature, and shown to be within the acceptable range of relative humidity.
Measurements are performed in a low-scatter environment to minimize increased
signal due to radiation scatter from the room walls, floor, and ceiling (thus
matching the chamber calibration conditions). Special calibration formalisms
are available for combinations of brachytherapy sources and treatment
applicators, electronic brachytherapy sources, LDR ophthalmic plaques
containing 106Ru/106Rh, and 90Y microspheres
used for selective internal radiotherapy of hepatocellular carcinoma or liver
metastases. Future brachytherapy calibration standards may employ the absorbed
dose-rate to water at 1 cm as it may provide lower propagated uncertainties
than the combination of RAKR (or SK) and the dose-rate constant.
However, such calibrations are not widely available for all sources, the gains
realized by lower propagated uncertainties are minimal, and clinical
brachytherapy treatment planning systems do not yet permit entry of this
quantity.